Precision Prep · CMD Exam Prep
Domain IV · Sheet 1 of 9

Photon Beam MU Calculations

Domain IV: Dose Calculation Methods
Exam Weight ≈ 13%
Key Formulas
SSD Technique: Monitor Units
MU  = 
Dose
OF × (PDD/100) × ISF × WF × TF
OF = Output Factor  |  PDD = Percent Depth Dose  |  ISF = Inverse Square Factor  |  WF = Wedge Factor  |  TF = Tray Factor
SAD Technique: Monitor Units
MU  = 
Dose
OF × TMR × WF × TF
TMR replaces PDD when isocenter is placed at the prescription depth.
No separate ISF needed: output is already referenced to the isocenter distance (SAD).
Inverse Square Factor (ISF)
ISF  = 
(SSDcal + dmax)²
(SSDtx + dmax)²
Corrects output when the treatment SSD differs from the calibration SSD.
ISF > 1 when treating at a shorter SSD (more output); < 1 when treating at a longer SSD.
Equivalent Square Field Size
S  = 
2 × a × b
a + b
Converts a rectangular field (a × b cm) to a square equivalent for output factor lookup.
Also expressed as: S = 4A / P where A = area and P = perimeter
Mayneord F-Factor: PDD Correction for SSD Change
F  = 
(SSD₂ + dmax)²
(SSD₁ + dmax)²
 × 
(SSD₁ + d)²
(SSD₂ + d)²
PDD₂ = PDD₁ × F  |  Corrects tabulated PDD when the treatment SSD differs from the table SSD
Quick Reference
Cal. SSD (typical)
100 cm
SAD (linac)
100 cm
6 MV dmax
1.5 cm
10 MV dmax
2.5 cm
15 MV dmax
3.0 cm
Co-60 dmax
0.5 cm
Memory Aids
SSD vs SAD: The Core Difference
SSD: source-to-skin distance is fixed; dose is prescribed at a depth from the surface. SAD: isocenter sits at depth; output is referenced to the isocenter. SAD requires TMR, not PDD.
ISF: Closer Means More Output
Treating closer than calibration distance → more dose per MU → ISF > 1 in the denominator → divide by a larger number → fewer MU needed.
Eq. Square Quick Check
10 × 20 field: S = (2×10×20)/(10+20) = 400/30 = 13.3 cm. Always less than the longer dimension and greater than the shorter one.
Field Modifiers Reference
ModifierSymbolTypical RangeEffect on MUNotes
Output FactorOF / Sc,p0.85 – 1.10MU ÷ OFIncreases with larger field size; largest at ~40 cm
Wedge FactorWF0.30 – 0.85MU ÷ WFLower WF = more MU; 60° wedge has lower WF than 15°
Tray / Block FactorTF0.95 – 0.97MU ÷ TFAccounts for tray attenuation (~3–5%)
Bolus----Shifts dmax to surfaceNo separate correction factor; adjusts effective depth
Exam Traps
SSD vs SAD: wrong formula
Using PDD in an SAD calculation is one of the most common errors. If the isocenter is placed at depth (SAD setup), use TMR. PDD applies when dose is measured at depth from the skin surface in an SSD setup.
Equivalent square uses harmonic mean
Eq. square = 2ab/(a+b), not the geometric mean √(a×b). For a 5 × 20 field: eq. sq. = 200/25 = 8.0 cm. Using the geometric mean gives 10 cm, too high, and leads to an incorrect output factor.
ISF uses dmax, not treatment depth
The ISF formula uses (SSD + dmax)² in both numerator and denominator: not (SSD + d). This is the source-to-reference-point distance, not the source-to-treatment-depth distance.
Mayneord is an approximation
The Mayneord F-factor works well for moderate SSD changes and shallow-to-mid depths. Accuracy decreases at large depths or very large SSD differences. Measured PDD tables at the actual SSD are more reliable.
Small fields → lower output
Output factor decreases with smaller field size because less phantom scatter reaches the detector. Small fields also have significant lateral electron disequilibrium. Do not assume output is 1.0 for small fields.
Bolus has no MU factor
Bolus is a tissue-equivalent material placed on the skin to reduce effective dmax. It does not have its own correction factor in the MU formula: the dosimetrist accounts for it through the adjusted depth used in the calculation.
Key Concepts to Know Cold
SSD setup uses PDD + ISF in the MU formula; SAD setup uses TMR with no ISF needed
Equivalent square: S = 2ab / (a+b); always use this for output factor lookup from square-field tables
ISF = (SSDcal + dmax)² / (SSDtx + dmax)²: corrects for change in source-to-reference distance
All modifiers (OF, WF, TF) sit in the denominator: lower values mean more MU are needed
Mayneord F-factor corrects PDD for a new SSD: PDD₂ = PDD₁ × F
dmax values: 6 MV = 1.5 cm  |  10 MV = 2.5 cm  |  15–18 MV = 3.0 cm  |  Co-60 = 0.5 cm
Wedge factor range: 0.30–0.85; a steeper wedge angle gives a lower WF and requires more MU
Tray factor ≈ 0.96: small but required whenever a block tray is present in the beam